WT37 Series Plastic Scintillator Detectors
for Measuring Alpha and Beta
TECHNICAL SPECIFICATIONS
  • WT370
  • WT372
  • WT375

ALPHA DETECTOR WT370 series


INTRODUCTION

WT370 Alpha Scintillator is a detector designed for alpha radiation survey when used with a general

The detector housing is made of aluminum alloy. The scintillator used to detect alpha rays is a piece of WT104 with a thickness of 1-1.5mm.

The ester film covers the surface of the scintillator. The role of the polyester film is to avoid light, but it needs to be very thin, otherwise the alpha rays will be partially or completely blocked. WT370 series detectors can be made into standard sizes, such as 2 inches, 3 inches, 5 inches, etc., and can also be made into other sizes according to user requirements. There is a voltage divider unit inside the detector, which outputs the original pulse signal.

The window of the detector is equipped with a cover to protect the polyester film of the detector from accidental damage.



Detector

WT104

Radiation detected

Alpha

Window

typically 1 mg/cm² aluminized Mylar

Communication

Original pulse

Mechanical

Aluminum metal housing

Operating voltage

Typically 700 to 1500 V DC

Diameter

Varied with model

Length

Varied with model

Weight

Varied with model

Operation temperature

-2050

Operation humidity

0 95% non-condensing


GET CONTACT WITH US
  • Name*

    *
  • Email*

    *
  • Phone*

    *
  • Message

    *
contact us
Whatsapp +86-15010461677
Email jean@wt-rad.com
Wechat
15010461677